19, Jalan PJS 3/59,
Taman Sri Manja,
46000 Petaling Jaya,
Selangor, Malaysia.
Previous | 44 / 54 | Next |
The Fluke i400E AC Current Clamp is an ideal tool to extend current measurement. It is designed to work with the new Fluke 101, 106 and 107 palm sized digital multimeters, as well as most Fluke multimeters or other voltage measurement tools that accepts 19.05 mm shrouded banana plugs and has mV ac measurement functionality
Fluke 106 Palm-sized Digital Multimeter
Fluke i400E AC Current Clamp
Accuracy is specified for 1 year after calibration, at operating temperatures of 18 °C to 28 °C, relative humidity at 0 % to 75 %. Accuracy specifications take the form of: ±([% of Reading] + [Number of Least Significant Digits])
Accuracy Specifications | ||||
Function | Range | Resolution | Accuracy | |
AC Volts1 |
6.000 V 60.00 V 600.0 V |
0.001 V 0.01 V 0.1 V |
1.0 % + 3 | |
DC Volts |
6.000 V 60.00 V 600.0 V |
0.001 V 0.01 V 0.1 V |
0.5 % + 3 | |
AC Millivolts | 600.0 mV | 0.1 mV | 3.0 % + 3 | |
Diode Test2 | 2.000 V | 0.001 V | 10 % | |
Resistance (Ohms) |
400.0 O 4.000 kO 40.00 kO 400.0 kO 4.000 MO 40.00 MO |
0.1 O 0.001 kO 0.01 kO 0.1 kO 0.001 MO 0.01 MO |
0.5 % + 3 0.5 % + 2 0.5 % + 2 0.5 % + 2 0.5 % + 2 1.5% + 3 |
|
Capacitance3 |
50.00 nF 500.0 nF 5.000 µF 50.00 µF 500.0 µF 1000 µF |
0.01 nF 0.1 nF 0.001 µF 0.01 µF 0.1 µF 1 µF |
2 % + 5 2 % + 5 5 % + 5 5 % + 5 5 % + 5 5 % + 5 |
|
Frequency4 Hz (10 Hz – 100 kHz) |
50.00 Hz 500.0 Hz 5.000 kHz 50.00 kHz 100.0 kHz |
0.01 Hz 0.1 Hz 0.001 kHz 0.01 kHz 0.1 kHz |
NA | |
Duty Cycle4 | 1 % to 99 % | 0.1 % | NA | |
AC Current (40 Hz to 200 Hz) |
4.000 A 10.00 A |
0.001 A 0.01 A |
1.5 % + 3 | |
DC Current |
4.000 A 10.00 A |
0.001 A 0.01 A |
1.5 % + 3 | |
|
||||
Function | Overload Protection | Input Impedance (Nominal) | Common Mode Rejection Ratio | Normal Mode Rejection Ratio |
AC Volts | 600 V 1 | >10 MO <100 pF 2 |
>60 dB at dc, 50 Hz or 60 Hz |
- |
AC Millivolts | 600 mV | >1M, <100 pF | >80 dB at 50 Hz or 60 Hz | - |
DC Volts | 600 V 1 | >10 MO <100 pF |
>100 dB at dc, 50 Hz or 60 Hz |
>60 dB at 50 Hz or 60 Hz |
|
||||
General Specifications | ||||
Maximum Voltage Between any Terminal and Earth Ground | 600 V | |||
Display (LCD) | 6000 counts, updates 3/sec | |||
Battery Type | 2 AAA, NEDA 24A, IEC LR03 | |||
Battery Life | 200 hours minimum | |||
Temperature | ||||
Operating | 0 °C to 40 °C | |||
Storage | -30 °C to 60 °C | |||
Relative Humidity | ||||
Operating Humidity |
Non-condensing when <10°C =90 % at 10 °C to 30 °C; =75 % at 30 °C to 40 °C |
|||
Operating Humidity, 40 MO Range | =80 % at 10 °C to 30 °C; =70 % at 30 °C to 40 °C | |||
Altitude | ||||
Operating | 2000 m | |||
Storage | 12,000 m | |||
Temperature Coefficient | 0.1 X (specified accuracy) /°C (<18 °C or >28 °C) | |||
Fuse Protection for Current Inputs |
11A, 1000V Fast Fuse, Fluke specified part only
|
|||
Size (HxWxL) | 142 mm x 69 mm x 28mm | |||
Weight | 200 g | |||
IP Rating | IEC 60529: IP 40 | |||
Safety | IEC 61010-1: 600 V CAT III, Pollution Degree 2 | |||
Electromagnetic Environment | IEC 61326-1: Portable | |||
Electromagnetic Compatibility | Applies to use the Korea only. Class A Equipment (Industrial Broadcasting and Communication Equipment)1 | |||
|
Electrical specifications | |
AC current range |
1 A to 400 A |
Output signal 1 | mV/A ac |
Accuracy |
2.0 % ± 5 digits (45 Hz to 65 Hz) 2.5 % ± 5 digits (65 Hz to 400 Hz) |
Typical bandwith | 5 Hz to 20 kHz |
Working voltage | 600 V ac rms, in compliance with IEC 61010-1 and IEC 61010-2-030 |
Load impedance | 1 A to 400 A |
Maximum non-destructive current | 600 A |
Jaw inner diameter | 30 mm |
Maximum wire size | 600 MCM |
Standard plugs | 19.04 mm dual banana plugs |
Safety specifications | |
Category rating |
IEC 61010-1 and IEC 61010-2-030/IEC 61010-2-031 and IEC 61010-2-032: 600 V CAT III Pollution Degree 2 |
Electromagnetic environment | IEC 61326-1: Portable |
General Specifications | |
Output cable length | 1 m |
Operating temperature | -10 °C to +50 °C |
Storage temperature | - 30 ºC to +60 ºC |
Relative humidity |
=90 % at 10 °C to 530 °C; =75 % at 30 °C to 540 °C; =45 % at 40 °C to 50 °C |
IP rating | IEC 60529: IP30 when jaw closed |
Temperature coefficients | Add 0.1 x specified accuracy for each °C above 28°C or below 18 °C. |
Operating altitude | 2,000 meters |
Storage altitude | 12,000 meters |
Vibration requirements | MIL-PRF-28800F, Class 2 |
Drop test requirements | -10 °C, 1m, 6 sides, oak wood |
Jaw impact test requirements | Follow EN/IEC 61010-2-032:2012 |
EMI, RFI, EMC | IEC/EN 61326-1 |
Dimensions (hxwxd) | 146 mm x 62 mm x 34 mm |
Weight | 150 g |
|